LED troubleshooting table
Section 9 Troubleshooting
40
AA003700-01 rev. G. 2.0
.
LED
State
Possible cause
Solution
Power
OFF
Power switch not ON
Turn power switch ON. Refer to
Connector Location
on page 26.
Power cable not connected
Make sure the power cable is
connected in the Spectrometer and in
the wall outlet. Refer to
Connector
Location
on page 26
No current at Power outlet
Test outlet for voltage.
Internal problems
Contact ABB. Refer to the back cover
of this document
GREEN
Normal operation
Status
RED
blinking
No detector connected
Replace or install detector module.
Refer to
Detector Module
Replacement
on page 29.
GREEN
with Red
short blinks
Operation Error
Check software health monitoring.
RED
Internal Error
Check software health monitoring.
OFF
System is loading
If after 1 min. LED is still OFF, check
software health monitoring
GREEN
Blinking
System is initializing or
performing calibration.
The initialization sequence is
performed at startup or after a
perturbation.
If this sequence last longer than 5
minutes, it is probably caused by a
low laser or white light level
Contact ABB. Refer to the back cover
of this document.
Table- 9- 1. LED Troubleshooting table