Section 7 Module Replacement
Source Module Replacement
AA003700-01 rev. G. 2.0
31
Source Module Replacement
Removing the Source Module
STEP 1
Power OFF the spectrometer. Refer to
Lockout/Tag Out Procedure
on page 10.
Class 3B invisible laser radiations (760 nm, 2mW output power, Near-IR) are
present in the interferometer module. However, no laser radiations leak out of the
Interferometer module unless the Source module is removed. To remove the
Source module refer to
Removing the Source Module
on page 31 and make sure
to use appropriate eye protection otherwise injuries will occur.
Use of controls or adjustments, or performance of procedures other than those
specified herein may result in hazardous radiation exposure.
Source Module is Hot! Always manipulate the Source module with caution: use
the non metallic parts to hold the Source module.
Ensure that equipment and any devices or power cords connected to it are
properly grounded.
The grounding pin of the power connect must be present at all times. If
necessary, have a certified electrician install a grounded wall outlet.
Protective earthing (grounding) must be active at all times. The absence of
grounding can lead to a potential shock hazard that could result in serious
injuries. If an interruption of the protective earthing is suspected, ensure the
equipment remains disconnected.
Always make sure the spectrometer is disconnected from a power source and
fully assembled (front and back panel installed) BEFORE opening, adjusting or
performing any maintenance.
Always use the Lockout/tag out procedure to power OFF the spectrometer. Refer
to
Lockout/Tag Out Procedure
on page 10.
Perform maintenance procedures in an ESD protected environment. Always use
an ESD protection such as wrist strap.