Direct Connection
Section 4 Connections
AA003700-01 rev. G. 2.0
20
Direct Connection
In this case, both the computer and the spectrometer will use a pre-configured LAN
configuration parameters.
If, when first turned ON, the spectrometer can not detect a DHCP server, it will
automatically use the following IP address:
10.127.127.127
The PC Ethernet interface should be configured to use a compatible set of LAN
parameters, for example:
IP address: 10.127.127.1
Subnet Mask: 255.255.255.0
To ensure your PC Ethernet interface is properly set, perform the following
procedure:
STEP 1
Open
Start> Settings> Network and Dial-up Connections.
STEP 2
In the
Network Dial-up connections
window, right click
Local Area
Connection
and select
Properties
. Refer to
Figure 4- 2
.
STEP 3
In the
Properties
dialog box, select
Internet Protocol (TCP/IP)
.
Figure 4- 4. Direct connection