8. Each step is displayed in the lower-right panel of the calibration screen as the
analyzer performs the calibration. Figure 42 shows the calibration process as a flow
chart.
Figure 42: Calibration Flow
9. When the
Calibration Complete message is displayed, click the CLOSE button.
10. Enter
TimeChart by selecting the Display button on the User Interface Control Bar,
and verify that the displayed concentration correctly corresponds to your local gas
standard.
11. Repeat steps 1-10 for each gas you want to calibrate.
The time of latest calibration is stored in
Reference Gas Settings within
the
Calibration menu for future reference.