5.6 Verifying the calibration
Introduction
Always verify the results after calibrating
any
robot axis to verify that all calibration
positions are correct.
Verifying the calibration
Use this procedure to verify the calibration result.
Note
Action
See
Run the calibration home position program twice.
Do not change the position of the robot axes after running
the program!
1
This is detailed in section
Synchronization marks and
synchronization position for
axes on page 341
Adjust the
synchronization marks
when the calibration is
done, if necessary.
2
Write down the values on a new label and stick it on top
of the calibration label.
3
xx
Remove any calibration equipment from the robot.
4
Product manual - IRB 1600/1660
367
3HAC026660-001 Revision: W
© Copyright 2006-2018 ABB. All rights reserved.
5 Calibration
5.6 Verifying the calibration
Содержание IRB 1660
Страница 1: ...ROBOTICS Product manual IRB 1600 1660 ...
Страница 2: ...Trace back information Workspace R18 2 version a18 Checked in 2018 11 20 Skribenta version 5 3 012 ...
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 48: ...This page is intentionally left blank ...
Страница 152: ...This page is intentionally left blank ...
Страница 336: ...This page is intentionally left blank ...
Страница 372: ...This page is intentionally left blank ...
Страница 388: ...This page is intentionally left blank ...
Страница 396: ......
Страница 397: ......