21
7
PROGRAMMING AND ELECTRICAL CALIBRATION…
…7.4 Set Up Outputs Page
Test Retrans (%)
Alter Sec. Code
0 0 0 0 0
ELECTRICAL CAL
– – – – –
0 0
Alter Cal. Code
0 0 0 0 0
Return to
top of page
.
Test Retransmission Output
The instrument automatically transmits a test signal of 0, 25, 50, 75 or 100% of the
retransmission range selected above. The % test signal selected is shown on the
upper display.
Example – for a selected range of 0 to 20 mA and 50% retransmission test signal,
10 mA is transmitted.
Select the required retransmission test signal.
Alter Security Code
Set the security code to a value between 00000 and 19999.
IMPORTANT – YOU
MUST
MEMORIZE THE NEWLY SET SECURITY CODE. If it
is forgotten contact the Company for advice.
Alter D.O. Calibration Code
Set the security code to a value between 00000 and 19999.
IMPORTANT – YOU
MUST
MEMORIZE THE NEWLY SET SECURITY CODE. If it
is forgotten contact the Company for advice.
Advance to Electrical Calibration Page, Section 7.7.
Continued from previous page.