12
4.1
Fitting the Dissolved Oxygen Sensor – Fig. 4.1
4
SETTING UP
Caution.
• Only install the oxygen sensor immediately prior to use, otherwise leave
it stored in its protective container.
• Take special care to line up the two pins in the oxygen sensor with their
respective sockets before making the connection and tightening.
• Take care not to damage the delicate membrane on the end of the
oxygen sensor.
• Ensure that the mating surfaces (carrying the electrical connection) of
the oxygen sensor and connector body are clean and completely dry.
Caution. Do not overtighten the clamping screw.
Fig. 4.1 Fitting the Dissolved Oxygen Sensor
Oxygen
Sensor
Clamping
Screw
Thrust
Washer
Flowcell
Connector
Body
Connector
Nut
3
Place an O-ring (provided) as shown
and locate the connector body
on the oxygen sensor.
Use the clamping screw to secure
the assembly. Screw in firmly using
finger pressure only.
7
5
Slide the thrust washer
over the connector body.
4
Slip the connector nut
over the connector body
and screw on to the
oxygen sensor firmly.
1
2
Remove the top from the oxygen sensor
container.
Unscrew the protective cap from the rear
of the oxygen sensor
6
I n s e r t t h e c o m p l e t e
assembly into the flowcell
ensuring that the O-ring is in
place.
O-ring
O-ring