FAM3200

 ARMORED VARIABLE AREA PURGEMETER  |  CI/FAM3200-EN REV. D 

11

 

  

 

 

8

 

Maintenance / Repair 

 

 WARNING 

Risk of injury due to process conditions. 

The process conditions, for example high pressures and 
temperatures, toxic and aggressive measuring media, can 
give rise to hazards when working on the device. 
•  Before working on the device, make sure that the process 

conditions do not pose any hazards. 

•  If necessary, wear suited personal protective equipment 

when working on the device. 

•  Depressurize and empty the device / piping, allow to cool 

and purge if necessary. 

 

 

 CAUTION 

Risk of burns due to hot measuring media 

The device surface temperature may exceed 70 °C (158 °F), 
depending on the measuring medium temperature! 
•  Before starting work on the device, make sure that it has 

cooled sufficiently. 

 

 

Note 
Loss of CE conformity! 

For pressure equipment consisting of assemblies, CE conformity 
is only for devices in the factory-supplied state. 
Components should only be replaced by the manufacturer's 
service personnel or an authorized specialist workshop. 
Replacing components yourself will render the CE conformity 
invalid. 
 

 

Spare parts 

 

Repair and maintenance activities may only be performed by 
authorized customer service personnel. 
When replacing or repairing individual components, use original 
spare parts. 
 

 

Note 

Spare parts can be ordered from ABB Service. 

www.abb.com/contacts

 

 

 

 

 

Cleaning 

 

Soiling of the meter tube and the float will impair the measuring 
accuracy of the device. The necessary cleaning interval is 
dependent on the operating conditions and must be set 
individually. 
The meter tube and the float must be removed in order to clean 
the device. 
 

 

When cleaning the exterior of meters, make sure that the 
cleaning agent used does not corrode the housing surface and 
the seals. 
To avoid static charge, a damp cloth must be used for cleaning. 
 

 

Disassembly of the meter tube 

 

NOTICE 

Damage to the float! 

Damage to the float due to incorrect disassembly. 
•  Observe the following points when disassembling the 

meter tube and the float! 

 

 
•  The float is precision made. During assembly/disassembly, 

ensure that the guide ring and the measuring edge are not 
damaged. A damaged float causes inaccuracies in the 
measurement and can cause damage to the meter tube 
under certain circumstances. 

•  Make sure that the meter tube is not exposed to impacts or 

mechanical loads during disassembly. 

 

FAM3220 / 3250 

FAM 3225 / 3255 

 

Figure 6:  Disassembly of the meter tube 

 

Содержание FAM3200 Series

Страница 1: ...le Area Purgemeter Measurement made easy FAM3200 Introduction Armored variable area purgemeters for measuring small flows of gases and liquids Additional Information Additional documentation on FAM320...

Страница 2: ...s 8 Pressure loss 8 Prevention of compression oscillations when measuring gases 8 Pressure shocks 8 Solids content in the measuring medium 9 Mounting 9 General information 9 Flowmeter installation 9 6...

Страница 3: ...he product may only be performed if expressly permitted by these instructions Information and symbols on the product must be observed These may not be removed and must be fully legible at all times Th...

Страница 4: ...measuring media may only be used if the operator is able to perform regular and suitable tests to ensure the safe condition of the device The operator bears sole responsibility for the use of the devi...

Страница 5: ...follower system 6 Cable entry Figure 1 Setup example The devices are used for local indication of the current flow rate with integrated alarm signalling units as flow rate monitors or with differenti...

Страница 6: ...he shipper without delay and before installation Transport The center of gravity of some devices is not at the center of the equipment Use any of the available attachment points on the device for tran...

Страница 7: ...and sudden pressure surges Use valves which open slowly If the flowmeter is installed in a pipeline where decommissioning is impossible or inexpedient a bypass line should be provided For gaseous mea...

Страница 8: ...ntion of compression oscillations when measuring gases During low flow amounts and low operating pressure so called compression oscillations of the float can occur If the maximum upstream pressure lis...

Страница 9: ...installation Install the flowmeter centrally in the piping at the required position using the correct threaded connections When tightening the process connections on the flowmeter use a spanner of the...

Страница 10: ...position transmitter in two wire technology Other cable lengths available on request 7 Commissioning CAUTION Risk of burns due to hot measuring media The device surface temperature may exceed 70 C 158...

Страница 11: ...s Repair and maintenance activities may only be performed by authorized customer service personnel When replacing or repairing individual components use original spare parts Note Spare parts can be or...

Страница 12: ...s conditions The process conditions for example high pressures and temperatures toxic and aggressive measuring media can give rise to hazards when dismantling the device If necessary wear suited perso...

Страница 13: ...Material load Figure 7 PS pressure TS temperature Maximum permissible operating pressure The permissible operating pressure is dependent on the optional device equipment Model Option Max permissible o...

Страница 14: ...0 7 10 15 1 0 14 5 8 0 12 04H 6 1 59 230 8 12 18 18 0 7 10 15 1 0 14 5 8 0 12 05H 10 2 64 350 12 36 18 18 0 7 10 15 1 0 14 5 8 0 12 06H 16 4 23 540 19 07 18 18 0 7 10 15 1 0 14 5 8 0 12 07H 25 6 6 850...

Страница 15: ...4 0 58 02 0 3 4 35 85 1 23 31H 600 158 5 18000 635 66 8 8 4 0 58 02 0 3 4 35 95 1 38 32H 700 184 92 21000 5547 61 8 8 4 0 58 02 0 3 4 35 105 1 53 33H 800 211 34 24000 847 55 8 8 4 0 58 02 0 3 4 35 13...

Страница 16: ...mentation declarations of conformity and certificates are available in ABB s download area www abb com flow Trademarks Buna N is a registered trademark of DuPont Dow Elastomers Kalrez and Kalrez Spect...

Страница 17: ...Company Address Contact person Telephone Fax Email Device details Type Serial no Reason for the return description of the defect Was this device used in conjunction with substances which pose a threa...

Страница 18: ...18 FAM3200 ARMORED VARIABLE AREA PURGEMETER CI FAM3200 EN REV D Notes...

Страница 19: ...FAM3200 ARMORED VARIABLE AREA PURGEMETER CI FAM3200 EN REV D 19 Notes...

Страница 20: ...ice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reser...

Отзывы: