![background image](http://html.mh-extra.com/html/abb/el3060-series/el3060-series_operating-instruction_4389109.webp)
EL3060
CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
109
Shewhart control charts
Shewhart control charts are used to determine the combined precision and
drift of the AMS.
The differences
d
t
between the measured values and the set points at the
zero point and the reference point are determined. These differences must
be less than or equal to the standard deviation
s
AMS
multiplied by a factor of
2. If the difference for a measured value is greater, the measured value is
marked with the status "Out of Control".