Password
Password protection
Access to the calibration and to those menus in which the configuration of
the instrument can be changed can be protected by a password. Password
protection is not activated ex works.
Password protection variants:
•
Access to the calibration menu is not included in the password protec-
tion.
•
Access to all instrument functions is included in the password protec-
tion (for instruments with SIL certification).
Configuring the password
The password is set in the configurator (see page 53) in the "Options –
Password…" menu. It consists of a four-digit number; each of the numbers
may only have the values 1, 2 and 3 (example: "1213"). The setting "0000"
means that password protection is not activated.