Insertion-Type Electromagnetic Probe Flowmeter
AquaProbe 2
5 Start Up And Operation
34
IM/AP
Issue 9
5.4 Replacing the Battery
Normal Operation
If both batteries are good, then no battery alarm is indicated.
Replace Battery
When a battery alarm is shown, replace the cell on the side
indicated. (In this example, the Right Battery).
Replace both batteries
If both batteries require replacement, it is important to first
change the cell indicated by the steady icon.
The flashing icon indicates the battery currently in use.
Proceed as follows:
1.
Remove the top cover of the transmitter – see section 3.1.3, page 19.
2.
Invert the cover.
3.
Slide out the connector from behind the battery retaining clamp and pull the connector apart.
4.
Remove the battery.
5.
Fit a new battery (ABB Limited Part No. WABC2001) ensuring that the end of the battery with the
connection wires is pushed up against the inside end of the top cover.
6.
Connect the battery connector to the connector previously removed.
7.
Push the connection centrally behind the battery retaining clamp to secure the battery.
8.
Fit the top cover to the transmitter and ensure the screws are tightened fully.
Caution.
If replacing the battery on an external powered (a.c. or d.c.) unit, ensure it is powered during
this operation. For dual battery units, replace only the battery above the indicated battery legend.
Note.
Each battery must be connected to the cable from the same side of the termination area as the
battery position in the lid.
6232709.
8
m
3
.9074
l / s
0
-
6232709.
8
m
3
.9074
l / s
0
-
6232709.
8
m
3
.9074
l / s
0
-
Содержание AquaProbe 2
Страница 1: ...Installation Guide IM AP Issue 9 Insertion Type Electromagnetic Probe Flowmeter AquaProbe 2 ...
Страница 40: ...Insertion Type Electromagnetic Probe Flowmeter AquaProbe 2 Notes 38 IM AP Issue 9 Notes ...
Страница 41: ...Insertion Type Electromagnetic Probe Flowmeter AquaProbe 2 Notes IM AP Issue 9 39 ...
Страница 42: ...Insertion Type Electromagnetic Probe Flowmeter AquaProbe 2 Notes 40 IM AP Issue 9 ...