Insertion-Type Electromagnetic Probe Flowmeter
AquaProbe 2
2 Mechanical Installation
IM/AP
Issue 9
13
2.6 Setting the Insertion Depth
2.6.1 Centre Line Method for Pipe Diameters
≤
1m (
≤
40in )
Warning.
When inserting or removing the AquaProbe suitable restraining equipment must be used to
prevent the probe being forced out under pressure. Ensure that the valve is fully open.
Note.
Safety restraint omitted for clarity.
Fig. 2.12 Setting the insertion Depth – Centre Line Method for Pipe Diameters 1m (40in)
1
2
9
Open Fully
Lower probe to
touch bottom
of pipe
Retract probe fully
Slide positioning collar
down to nut and lock
Slacken
Tighten
to 40Nm
(30ft lbf)
Determine
internal
diameter
(D)
Unlock, slide
positioning collar
down and lock at distance:
30mm (1.181in)
30mm (1.181in)
D
2
D
2
Insert probe
to position
collar
depth
8
7
3
5
6
4
See
Note