Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
6 Configuration
30
OI/AXM630–EN Rev. I
6.1.4 Security
The analyzer is fitted with an internal security switch that, in combination with the 'Configuration security' parameter settings (see
page 31), is used to prevent unauthorized access to the Configuration Level.
Two methods of configuration access protection are available:
1.
Password protection
(Factory Default)
The Configuration level can be accessed only when the correct password has been entered.
2.
Internal security switch protection
The Configuration level can be accessed only when the internal security switch is set to the 'Enabled' position.
To access the internal security switch:
1. Switch off the power supply to the analyzer and turn the two door retaining screws
A
1
/
4
turn counter-clockwise.
2. Using a cross-head screwdriver, remove the four cover plate retaining screws
B
and remove the cover plate.
3. Set the security switch
C
to the required position.
'Configuration security' parameter setting
(see page 31)
Internal Security Switch Setting
'Password protected'
(Factory Default)
'Internal switch protected'
(Alternative)
Disabled
(Factory Default)
Password Access
No Access
Enabled
Free Access
Free Access
Fig. 6.2 Accessing the Internal Security Switch
Note.
The Internal Security Switch is, by default, set to 'Disabled' and should only be used to access the Configuration level
when 'Configuration security' is set to 'Internal switch protected' – see page 31.
Do Not
use the switch to access the
Configuration level when 'Configuration security' is set to 'Password protected' (default setting) unless the Password has been
forgotten. The switch overrides Password protection, enabling free access to the Configuration level.
A
B
C
Switch '
Disabled
' Position:
Configuration Level Access Disabled
Switch '
Enabled
' Position:
Configuration Level Access Enabled
Enabled
Disabled
Disabled
Enabled