Appendix F – Updating the Software
OI/AXM630–EN Rev. I
89
Appendix F – Updating the Software
1. Stop analyzer operation by selecting 'Stop Monitor' from
the main operator menu.
2. Ensure the analyzer is switched off.
3. Use a large flat-headed screwdriver to release the two
door catches
A
.
4. Remove the SD card
B
.
5. Refer to the documentation on the web site
for details of any software updates.
6. Download the software and transfer it to a blank SD card
using an appropriate card reader.
7. Set the switch
C
(above the SD card socket) to the lower
position (Software Update).
8. Insert the SD card, close the door and switch on the
analyzer. A start-up message appears for a short duration
(approximately 5 seconds) followed by a 'File Load
Progress' bar then the 'ABB' splash screen:
9. A 'Warning' screen appears next. This provides the option
to upgrade the existing system software or exit without
upgrading:
10. To proceed, press the
key.
11. The 'Commissioning V2.0' screen appears with the
'Upgrade Firmware' button selected:
12. To upgrade the existing system software, press the
key to start the process. The software is transferred to the
analyzer in sections.
13. Press the
key as each section is installed (installation
may take a few minutes).
14. Open the door and move the switch to the up position.
15. Close the door and re-tighten the door catches (A in Fig.
F.1).
16. Turn power off and then on to restart the analyzer.
Fig. F.1 SD Card Removal
A
B
C
Normal
Operation (Up)
Software
Update
(Down)