Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
6 Configuration
OI/AXM630–EN Rev. I
31
Note.
User 1 is the
System Administrator
and the only user with access to the 'Security type' parameter – see Table 6.1.
User 1
User 1 (System Administrator) Security Rights
System Administrator
Set initial password-protected access to Calibration & Maintenance and Logging menus.
Is the only user with access to the 'Security type' parameter.
Set initial password-protected access to the Configuration menu when 'Security
type / Configuration security' parameter is set to 'Password Protected'.
Set initial user permissions – other users can subsequently change their own passwords if
permission has been set by User 1.
Set password expiry dates and disable Inactive user accounts after a set time.
Set password failure limits and minimum password lengths
Table 6.1 System Administrator Security Rights
Fields
Description
Security type
A page opens with two fields:
Security system
– toggles between 'Basic' and 'Advanced'.
–
Basic:
Allows access to the 'Configuration' menu for up to four users (User 1 to 4) – a unique
password of up to four digits can be set for each user.
A separate password can be set to access the 'Calibration & Maintenance' and
'Logging' menus – up to four users share this password.
–
Advanced:
Allows up to twelve users password-protected access to any of the 'Configuration',
'Calibration & Maintenance' or 'Logging' menus.
Each user can be assigned a unique 20-digit (alphanumeric) case-sensitive password
(minimum password lengths can be set).
Configuration security
– toggles between 'Password protected' and 'Internal switch
protected'.
–
Password protected (factory default):
With the internal security switch set to 'Disabled' (factory default), the 'Configuration'
level can be accessed only when the correct password is entered.
–
Internal switch protected:
With the internal security switch set to 'Disabled' (factory default), the 'Configuration'
level cannot be accessed.
Note.
If the internal security switch is set to 'Enabled', the Configuration Level can be accessed
without the need to enter a password and should be used only if the password has been
forgotten.