32
Art: 723085-00K Rev. Date: 30-OCT-2014
Deleting Results From Analyzer
The Utility Function is used to delete stored results. This function can be password
protected using the Customization function.
To delete stored results:
1. Press to turn on analyzer
2. Press
to change screen to Administration Menu
3. Press (Utility)
4. When prompted for Password, press the
Enter key or the enter the password.
5. Press (Clear Memory)
6. Select from the Clear Memory menu:
1. Previous to current month and year
(as shown)
2. Previous to previous month and year
(as shown)
3. All
4. Cancel
2
7
MENU
15:26 18JUNyy
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Utility
1 – Send Software
2 – Clear Memory
3 – Receive Software
Utility
Clear Memory
1 – Previous to
01JUL09
2 – Previous to
01JUN09
3 – All
4 – Cancel