34 User Guide for the Platinum Series Sound Processor
User Guide for the Platinum Series Sound Processor 35
Guidance and Manufacturer’s Declaration
Per IEC 60601-1-2
Table 1: Guidance and Manufacturer’s Declaration-Electromagnetic emissions
X-Ray Machines
X-Ray machines will not damage the implant but may damage the headpiece
microphone . Avoid placing the headpiece in checked or carry-on baggage that is
screened with x-rays . During airport security screening, the headpiece should either
be worn through the metal detector or be hand-examined .
Recommended System Operating and Storage Temperatures
Table: Operating and storage temperature ranges
Condition
Minimum
Maximum
Operating temperature
0°C (32°F)
45°C (115°F)
Storage temperature
-20°C (-4°F)
55°C (131°F)
The Platinum Series system meets the temperature safety requirements of IEC
60601-1:2005 in ambient temperatures up to 33˚C .
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration –electromagnetic emissions
The Platinum Series system is intended for use in the electromagnetic environment
specified below . The customer or the user of the Platinum Series system
should assure that it is used in such an environment .
Emissions test
Compliance
Electromagnetic environment - guidance
The Platinum Series system uses RF
energy only for its internal function .
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment .
Group 1
Class B
Not applicable
Not applicable
The Platinum Series system is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes .