Page
14
AD-4405 Weighing Indicator
Specifyi
Specifyi
Specifyi
Specifying
ng
ng
ng the Weighing Range of the First Range
the Weighing Range of the First Range
the Weighing Range of the First Range
the Weighing Range of the First Range
Step 3 After displaying
Cap
for 2 seconds, the first range or the weighing capacity of
single range will be displayed. When dual range is used,
Cap1
is displayed for 2
seconds. Triangle 1 will be displayed. Specify parameter with the
0
-
9
keys
and press the
ENTER
key to store it and proceed to the next step. When pressing
the
ESC
key, regardless of what is displayed, the indicator will proceed to the next
step without changing the parameters. The next step is Zero Calibration in single
range or the second range resolution in dual range.
Specifying
Specifying
Specifying
Specifying the Second Range Resolution
the Second Range Resolution
the Second Range Resolution
the Second Range Resolution
Step 4 After displaying
range2
for 2 seconds, the resolution with decimal point and triangle
2 will be displayed. Specify the second range resolution in the same way as the
first range. The decimal point cannot be moved. Specify the second range resolution
greater than the first range. Press the
ENTER
key to store the parameters and
proceed to the next step. When pressing the
ESC
key, regardless of what is
displayed, the indicator will proceed to the next step without changing the parameter.
Specifying
Specifying
Specifying
Specifying the Second Range Capacity
the Second Range Capacity
the Second Range Capacity
the Second Range Capacity
Step 5 After displaying
Cap2
for 2 seconds, the capacity with unit and decimal point is
displayed. Specify the capacity in the same way as the first range. The capacity
should be greater than the first range. Press the
ENTER
key to store the parameters
and proceed to Zero Calibration.
5.2.2.
To Get Stabilized Data
Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.
Maintain a constant temperature, stable power and stable input voltage from the
load cell.
Avoid direct sunshine or the near the outlet of an air conditioner.
Do not install the scale (indicator) where there is a strong magnetic field.
Step 7 Turn the display on and leave it for several minutes.
5.2. Calibration Procedure
Содержание Weighing Indicator AD-4405
Страница 1: ...Weighing Indicator WM PD4000504...
Страница 76: ...Page 74 AD 4405 Weighing Indicator MEMO...
Страница 77: ...AD 4405 Weighing Indicator Page 75 MEMO...
Страница 78: ...Page 76 AD 4405 Weighing Indicator MEMO...
Страница 79: ...13...
Страница 80: ......