EMD Technical Data Battery-operated Blood Pressure Monitor 

Medical Electrical Equipment needs special precautions regarding EMD and needs to be 
installed and put into service according to the EMD information provided in the following. 
Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical 
Electrical Equipment. 
The use of accessories and cables other than those specified may result in increased 
emissions or decreased immunity of the unit. 

 
Table 1 

EMISSION Limits

 

Phenomenon 

Compliance 

Conducted and radiated RF EMISSION

   

   

 

CISPR 11

Group 1

Class B 

 
Table 2 

IMMUNITY TEST LEVELS

Enclosure Port

 

Phenomenon 

IMMUNITY TEST LEVELS 

Electrostatic discharge 

 

IEC 61000-4-2 

±8 kV contact   
±2 kV, ±4 kV, ±8 kV, ±15 kV air 

Radiated RF EM fields 

 

IEC 61000-4-3 

10 V/m 
80 MHz - 2.7 GHz 
80 % AM at 1 kHz 

Proximity fields from RF wireless communications equipment 

IEC 61000-4-3 

See table 3 

Rated power frequency magnetic fields 

 

IEC 61000-4-8 

30 A/m 
50 Hz or 60 Hz 

 
Table  3 

Test  specifications  for  ENCLOSURE  PORT  IMMUNITY  to  RF  wireless  communications 

equipment

 

Test 

frequency 

(MHz) 

Band 

(MHz) 

Service 

Modulation 

Maximum 

power 

(W) 

Distance 

(m) 

IMMUNITY TEST 

LEVEL 

(V/m) 

385 

380 

 390 

TETRA 400 

Pulse modulation 

18 Hz 

1.8 

0.3 

27 

450 

430 

 470 

GMRS 460 

FRS 460 

FM 

±5 kHz deviation 

1 kHz sine 

0.3 

28 

710 

704 

 787 

LTE Band 13,17 

Pulse modulation 

217 Hz 

0.2 

0.3 

745 
780 

810 

800 

 960 

GSM 800/900 

TETRA 800 

iDEN 820 

CDMA 850 

LTE Band 5 

Pulse modulation 

18 Hz 

0.3 

28 

870 

930 

1720 

1700 

 1990 

GSM 1800 

CDMA 1900 

GSM 1900 

DECT 

LTE Band 1,3,4,25 

UMTS 

Pulse modulation 

217 Hz 

0.3 

28 

1845 

1970 

2450 

2400 

 2570 

Bluetooth 

WLAN 802.11 b/g/n 

RFID 2450 

LTE Band 7 

Pulse modulation 

217 Hz 

0.3 

28 

5240 

5100 

 5800 

WLAN 802.11 a/n 

Pulse modulation 

217 Hz 

0.2 

0.3 

5500 
5785 

 

Содержание UM-102

Страница 1: ...Instruction Manual Original Manuel d instructions Traduction Manual de instrucciones Traducci n Manuale di Istruzioni Traduzione 1WMPD4002909B...

Страница 2: ......

Страница 3: ...ction Manual Contents 1 Dear Customers 2 2 Preliminary Remarks 3 3 Notes for Proper Use 4 4 Parts Identification 7 5 Symbols 8 6 Using the Monitor 9 7 Measurements 12 8 Useful Features 13 9 Troublesho...

Страница 4: ...se The device is designed for use in a medical facility Intended purpose This device is designed to measure human blood pressure and heart rates Features Measurement This sphygmomanometer is designed...

Страница 5: ...parately if the device will not be used for a month or more Otherwise the batteries may degrade The battery life varies with the ambient temperature and may be shorter at low temperatures Generally tw...

Страница 6: ...rated oxygen environments such as a high pressure oxygen chamber or an oxygen tent It may cause a fire or explosion When the cuff is infected by blood or body fluid it should be safely disposed of acc...

Страница 7: ...through the artery This may cause pain numbness or a temporary red mark to the arm This condition will appear especially when measurement is repeated successively Any pain numbness or red marks will...

Страница 8: ...ter exercising or taking a bath A patient must rest for twenty to thirty minutes before the measurement Notes for Proper Measurement Let a patient sit down in a comfortable position Confirm that the p...

Страница 9: ...merical display During measurement After exhausted air Indicates pressure value Indicates pulse rate Display Unit of pressure Pressure value Battery indicator Pulse rate Unit of pulse High digits of t...

Страница 10: ...s Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Pressure remains in the cuff E...

Страница 11: ...ries It may shorten the battery life or cause the device to malfunction Battery life varies with the ambient temperature and may be shorter at low temperatures Only the specified batteries should be u...

Страница 12: ...cuff CUF KW LL 31 cm to 45 cm Large Adult cuff CUF KW LA 22 cm to 32 cm Adult cuff CUF KW A 16 cm to 24 cm Small Adult cuff CUF KW SA 12 cm to 17 cm SS cuff CUF KW SS Arm size The circumference of th...

Страница 13: ...nge ARTERY Artery position mark THIS SIDE TO PATIENT Instructions to the patient REF Catalog number Lot number ADULT Cuff size 22 cm to 32 cm LL Cuff size 41 cm to 50 cm LARGE ADULT Cuff size 31 cm to...

Страница 14: ...ial artery and pressurize the cuff by squeezing the rubber bulb Make sure the exhaust valve is completely closed While the cuff is inflating the pressure bar will move and in turn the LCD will display...

Страница 15: ...ction Allow at least three minutes between measurements on the same person 8 Useful Features Measurement with MARK Button You can put a marker at a certain pressure value when the MARK button is press...

Страница 16: ...tact your nearest authorized A D dealer Do not attempt to open or repair this product by yourself as any attempt to do so will render your warranty invalid 10 Maintenance Do not attempt to open the de...

Страница 17: ...enzalkonium Chloride Solution 0 05 Sodium Hypochlorite 0 05 Do not wipe off the valve and its vicinity with gauze soaked in disinfectant solution It will cause stuffing of the rubber bulb Clean the de...

Страница 18: ...to 40 C 15 RH to 85 RH 800 hPa to 1060 hPa Transport Storage conditions 20 C to 60 C 10 RH to 95 RH 700 hPa to 1060 hPa Dimensions Approx 98 W x 324 H x 67 D mm Weight Approx 520 g excluding batteries...

Страница 19: ...10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 See table 3 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Tabl...

Страница 20: ...English 18 MEMO...

Страница 21: ...lifornia 95131 U S A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D RUSO...

Отзывы: