English      13 

6.  Turn  the  exhaust  valve  screw  counterclockwise  to 

release all the air from the cuff. 

  If a measurement is taken with insufficient pressure, 

the 

  mark will be displayed. Re-pressurize the 

cuff to a pressure that is about 30 to 40 mmHg 
higher than the previous attempt. 

  An error message   

    will be displayed if a 

measurement is taken with insufficient pulses or in 
a very noisy environment. 

 
7.  Press  the 

    START/STOP

  button  again  to  turn  off  the 

power. 

 
Note:  Model UM-102 has an automatic power-off function. 

Allow at least three minutes between measurements on the same person. 

 
 

8.

 

Useful Features 

Measurement with MARK Button   

 

You can put a marker at a certain pressure value 
when  the 

  MARK 

  button  is  pressed  during  the 

measurement process. Up to 5 markings can be 
shown over the range of 40mmHg. 

 

Lifetime Counter 

 

When  the 

  MARK 

  button  is  pressed  while  the 

device is off, the lifetime counter is displayed.   
This  counter  function  indicates  the  hours  the 
device was in use and helps to determine when 
maintenance  is  necessary.  High  digits  and  Low 
digits  are  alternatively  displayed.  The  example 
indicates  that  the  device  has  been  in  use  for 
1,278 hours. 

 
 

Marker 

Marker 

START/STOP

High digit 

Low digit 

Alternatively

Содержание UM-102

Страница 1: ...Instruction Manual Original Manuel d instructions Traduction Manual de instrucciones Traducci n Manuale di Istruzioni Traduzione 1WMPD4002909B...

Страница 2: ......

Страница 3: ...ction Manual Contents 1 Dear Customers 2 2 Preliminary Remarks 3 3 Notes for Proper Use 4 4 Parts Identification 7 5 Symbols 8 6 Using the Monitor 9 7 Measurements 12 8 Useful Features 13 9 Troublesho...

Страница 4: ...se The device is designed for use in a medical facility Intended purpose This device is designed to measure human blood pressure and heart rates Features Measurement This sphygmomanometer is designed...

Страница 5: ...parately if the device will not be used for a month or more Otherwise the batteries may degrade The battery life varies with the ambient temperature and may be shorter at low temperatures Generally tw...

Страница 6: ...rated oxygen environments such as a high pressure oxygen chamber or an oxygen tent It may cause a fire or explosion When the cuff is infected by blood or body fluid it should be safely disposed of acc...

Страница 7: ...through the artery This may cause pain numbness or a temporary red mark to the arm This condition will appear especially when measurement is repeated successively Any pain numbness or red marks will...

Страница 8: ...ter exercising or taking a bath A patient must rest for twenty to thirty minutes before the measurement Notes for Proper Measurement Let a patient sit down in a comfortable position Confirm that the p...

Страница 9: ...merical display During measurement After exhausted air Indicates pressure value Indicates pulse rate Display Unit of pressure Pressure value Battery indicator Pulse rate Unit of pulse High digits of t...

Страница 10: ...s Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Pressure remains in the cuff E...

Страница 11: ...ries It may shorten the battery life or cause the device to malfunction Battery life varies with the ambient temperature and may be shorter at low temperatures Only the specified batteries should be u...

Страница 12: ...cuff CUF KW LL 31 cm to 45 cm Large Adult cuff CUF KW LA 22 cm to 32 cm Adult cuff CUF KW A 16 cm to 24 cm Small Adult cuff CUF KW SA 12 cm to 17 cm SS cuff CUF KW SS Arm size The circumference of th...

Страница 13: ...nge ARTERY Artery position mark THIS SIDE TO PATIENT Instructions to the patient REF Catalog number Lot number ADULT Cuff size 22 cm to 32 cm LL Cuff size 41 cm to 50 cm LARGE ADULT Cuff size 31 cm to...

Страница 14: ...ial artery and pressurize the cuff by squeezing the rubber bulb Make sure the exhaust valve is completely closed While the cuff is inflating the pressure bar will move and in turn the LCD will display...

Страница 15: ...ction Allow at least three minutes between measurements on the same person 8 Useful Features Measurement with MARK Button You can put a marker at a certain pressure value when the MARK button is press...

Страница 16: ...tact your nearest authorized A D dealer Do not attempt to open or repair this product by yourself as any attempt to do so will render your warranty invalid 10 Maintenance Do not attempt to open the de...

Страница 17: ...enzalkonium Chloride Solution 0 05 Sodium Hypochlorite 0 05 Do not wipe off the valve and its vicinity with gauze soaked in disinfectant solution It will cause stuffing of the rubber bulb Clean the de...

Страница 18: ...to 40 C 15 RH to 85 RH 800 hPa to 1060 hPa Transport Storage conditions 20 C to 60 C 10 RH to 95 RH 700 hPa to 1060 hPa Dimensions Approx 98 W x 324 H x 67 D mm Weight Approx 520 g excluding batteries...

Страница 19: ...10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 See table 3 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Tabl...

Страница 20: ...English 18 MEMO...

Страница 21: ...lifornia 95131 U S A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D RUSO...

Отзывы: