HV-G/HW-G Series
Page 41
12. Full/Dribble Batch Function
Caution
The comparison of the full/dribble batch function can not be restored. (One
way sequence). If the weight value becomes less than the final value after the
value reached a predetermined target value, neither HI nor LO is on.
The parameters of the upper limit value (HI) and the final value (HI) use the
same memory. The parameters of the lower limit value (LO) and the
preliminary value (OK) use the same memory.
The upper/lower comparator function, the simple batch function and the
full/dribble batch function can not be used at the same time because these
parameters use the common memory.
Set the zero band greater than the tare value.
Operation
The batch process starts when the
SET
key is pressed.
When a parameter from
f10 0
or
f10 2
of the full/dribble batch sub-function is
selected, the F key functions as the finish key.
Parameter List and Word Definition
The "gross" is a total weight value (the tare value is not subtracted).
The "net" is a measurement value with the tare value subtracted from the gross.
The "zero band" means the zero detection level.
The "zero point" is the fundamental starting point to weigh anything.
Comparator
Function table
Description
A
f6 9
Full/dribble batch function.
A
Full/dribble batch sub-function
Function table
Description
A
f10 0
When the
F
key is pressed after final value is reached, the current
process is finished.
A
A
f10 1
When the stability mark is displayed after final value is reached,
the current process is finished.
A
A
f10 2
When the
SET
key is pressed, the scale automatically tares and
starts the full/dribble batch process. When the
F
key is pressed
after final value is reached, the current process is finished.
A
A
A
f10 3
When the
SET
key is pressed, the scale automatically tares and
starts the full/dribble batch process. When the stability mark is
displayed after final value is reached, the current process is
finished.
A
A
A
Hold
Function table
Description
A
f12 0
The hold function is not used.
A
Содержание HW-10KGL
Страница 19: ...HV G HW G Series Page 17 6 Description of Each Part 6 Description of Each Part...
Страница 20: ...Page 18 HV G HW G Series 6 Description of Each Part...
Страница 67: ...HV G HW G Series Page 65 17 Options For models with weighing capacity of 300 kg to 1200 kg...
Страница 81: ...HV G HW G Series Page 79 18 Specifications Dimensions Unit mm...
Страница 82: ...Page 80 HV G HW G Series 18 Specifications Unit mm...
Страница 84: ...Page 82 HV G HW G Series MEMO...
Страница 85: ...HV G HW G Series Page 83 MEMO...
Страница 86: ...Page 84 HV G HW G Series MEMO...