HV-G/HW-G Series
5. Installing the Scale
5.3. Removing
the
Pole
Caution
Remove the AC plug from the receptacle before removing the pole (type V).
Remove the AC adapter and batteries before removing the pole (type L).
When removing the load cell cable, do not pull on the load cell cable
connector forcibly and do not pull on the wires of the cable.
Do not bend the cable forcibly. Use care so that the load cell cable does not
touch the pan inside the base unit.
Avoid dust, static electricity and high humidity (or condensation) because the
inside of the display unit is sensitive.
The pole removing procedure depends on the scale model. Refer to the table below.
Models
Refer to
HV-15KGL/HV-15KGV/HV-60KGL/HV-60KGV/HV-200KGL/HV-200KGV/HW-10KGL
HW-10KGV/HW-60KGL/HW-60KGV/HW-100KGL/HW-100KGV/HW-200KGL/HW-200KGV
Procedure A
HW-300KGL4/HW-300KGV4/HW-600KGL4/HW-600KGV4/HW-600KGL3/HW-600KGV3
HW-1200KGL3/HW-1200KGV3
Procedure B
5.3.1. Procedure A
Step 1 Turn the scale off.
Remove the AC plug from the receptacle (type V).
Remove the AC adapter and batteries (type L).
Step 2 Open the rear cover of the display unit.
Disconnect the load cell cable connector gently
(perpendicularly and do not pull toward you).
Step 3 Remove the two 4 mm screws, which are used
to attach the display unit to the pole.
Step 4 Remove the ferrite core and the cable clamp
from the load cell cable.
Step 5
(HV-60KGL/HV-60KGV/HW-60KGL/HW-60KGV/HV-
200KGL/HV-200KGV/ /HW-100KGL/HW-100KGV
HW-200KGL/HW-200KGV only)
Remove the 3 mm screws from the bottom
cover of the bracket.
Step 6 Carefully remove the load cell cable from the
pole and the bracket while using care not to pull
on the connector forcibly.
Step 7 Arrange the cable so that it does not touch the
pan in the base unit. The untied cable is
approximately 2 m. The optional extension load
cell cable (OP-02) is 5 m long.
Step 8 Remove the bracket from the base unit. An
Allen wrench is required.
HV-15KGL/HV-15KGV
HV-60KGL/HV-60KGV
HW-10KGL/HW-10KGV
HW-60KGL/HW-60KGV
HV-200KGL/HV-200KGV
HW-100KGL/HW-100KGV
HW-200KGL/HW-200KGV
5 mm Allen wrench
6 mm Allen wrench
Bracket
Allen screw
Allen wrench
Bracket (Middle and large sized models)
Pole
Bottom cover
3 mm screws
Load cell cable
Rear cover
Cable clamp
Ferrite core
Wind up 2 times
4 mm screw
Connector
Содержание HW-10KGL
Страница 19: ...HV G HW G Series Page 17 6 Description of Each Part 6 Description of Each Part...
Страница 20: ...Page 18 HV G HW G Series 6 Description of Each Part...
Страница 67: ...HV G HW G Series Page 65 17 Options For models with weighing capacity of 300 kg to 1200 kg...
Страница 81: ...HV G HW G Series Page 79 18 Specifications Dimensions Unit mm...
Страница 82: ...Page 80 HV G HW G Series 18 Specifications Unit mm...
Страница 84: ...Page 82 HV G HW G Series MEMO...
Страница 85: ...HV G HW G Series Page 83 MEMO...
Страница 86: ...Page 84 HV G HW G Series MEMO...