Page 38
HV-G/HW-G Series
11. Upper/Lower Comparator Function
Step 5 If either of
f6 0
,
f6 2
,
f6 4
,
f6 6
has been
selected, press the
SET
key to use the
comparator.
Step 6 Press the
MODE
key to display the blinking HI.
Step 7 Set the upper limit value using the following keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
F
key
Selecting the polarity (+,-).
Step 8 Press the
ENTER
key to store the new parameter and
display the blinking LO.
Step 9 Set the lower limit value using the following keys.
<
key
Selecting the figure to be changed.
∧
key
Changing the number of the figure.
F
key
Selecting the polarity (+,-).
Step 10 Press the
ENTER
key to store the new parameter and
return to the normal weighing mode.
If the lower limit value is greater than the upper limit
value, the scale returns to step 7.