HW-G Instruction Manual Page 41
HW-G OZIM v1a
14.2 The Complete Calibration Procedure
14.2.1 Gravity Acceleration Correction
Step1
Turn on the display.
Open the rear cover of the display unit.
Inside the display unit, press and hold the
CAL
switch to enter the calibration mode.
Then
Cal 0
is displayed.
Step2
Press the
ZERO
switch to enter the gravity
acceleration correction mode.
Step3
Set your local gravity acceleration using
the following switches.
^
switch Selecting the value of a digit.
<
switch Selecting a digit.
Step4
Press the
ENTER
switch to store the new
value.
Step5
Press the
CAL
switch again. Then the
scale returns to the normal weighing
mode.
14.2.2 Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and a stable power supply.
Install the scale on a solid floor or bench where there is no draft,
vibration, strong magnetic fields or direct sunlight.
Consider section “4. Caution”.
Step 7 Display normal weighing for at least 30 minutes to warm up the scale.