14.2. Complete Calibration Procedure
14.2.1. Gravity Acceleration Correction
Step 1 Turn on the display.
Open the rear cover of the display unit. Locate the CAL
switch inside.
Press
the
CAL
switch to enter calibration mode.
Then
is displayed.
Step 2 Press the
MODE
key to enter gravity acceleration
correction mode. Press the
ENTER
key to make the
numerical value blink and proceed to input mode.
Step 3 Set your local gravity acceleration using the numerical keys.
Step 4 Press the
ENTER
key to store the new value.
The display returns to gravity acceleration display after
displaying
.
Step 5 Press the
CAL
switch again to finish gravity acceleration
correction.
14.2.2. Preparation
Step 1 Confirm the environmental conditions as follows:
Maintain a constant temperature and stable power.
Install the scale on a solid floor where there is no draft, vibration, strong magnetic fields or
direct sunlight.
Refer
to
"
3.1. Installing the Scale
".
Step 2 Keep the display turned on for at least 30 minutes to warm up the scale.
CAL switch
Use numerical keys
CAL
switch
The display is automatically turned off.