4
5. Place the
ionizer on the stand and set it up onthe side of the balance.
(Secure the ionizer to the stand using the double-sided tape on
the stand if necessary)
1.3. How to Use
Note
Keep a space between electrodes and the weighing sample. Placing the sample to close to an
electrode may cause the sample to become charged.
The electrodes must not be touched with tweezers, etc. while neutralizing.
1. Place the weighing sample to be neutralized in front of the ionizer.
At this time, the weighing sample to be neutralized must be kept in a range of approx. 10 cm to 30 cm from
the electrodes as shown in the figure below.
Neutralization ion generation distribution
Top view
Side view
Weighing sample
Electrodes
Ion
area
10cm 20cm 30cm 50cm
Effective
neutralization
area
Effective
neutralization
area
Ion
area
Stand
Double-sided tape