3. Maintenance of the Electrode Unit
In general, when using the ionizer for a long time, dust and
stains may stick to the electrodes. Clean them periodically
to maintain performance.
When pins of the electrode are rubbed down and the
neutralization function does not recover after cleaning
them, replace the electrodes with four new units. The
standard life time is approximately 10000 hours.
Replacing Electrode Unit
1
Turn the electrode unit 45 degrees counterclockwise. Remove it.
2
Replace all two units with new ones at the same time.
Note
Do not remove and clean the electrodes while the
mark and the LED are flashing.
4
.
Specifications
Static elimination method
DC corona discharge (Bipolar)
Elimination range
Approx. 10 cm to 30 cm from
the electrode
Elimination performance
1 sec. (at
a distance
of approx. 10 cm and 4kV electrical charge)
Ambient temperature and
humidity
5 to 45
℃
, 85
%
RH or less (No condensation)
Electrode
Tungsten (life time of approx. 10,000 hours)
Mass
Approx. 280 g (including the stand)
5
.
Option
Electrode Unit of the DC Static Eliminator AX-BM-NEEDLESET (4 pieces)
Use these electrode units for replacement parts of the
DC static eliminator built into the ionizer.
Replace all two units with new ones at the same time.
Refer to "3. Maintenance of the Electrode Unit " for the
replacement.
Electrode unit
Step 1
Step 2
Cleaning
Electrode