41
Zero tracking (
trc
)
This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When the weighing data is only a few digits, turn the function off for accurate weighing.
Note
Digit, when used for the GP series balance, indicates the smallest displayable weighing value.
trc 0
The tracking function is not used. Used for weighing a very light sample.
trc 1
The tracking function is used. Normal zero tracking.
trc 2
The tracking function is used. Strong zero tracking.
trc 3
The tracking function is used. Very strong zero tracking.
Display refresh rate (
5pd
)
Period to refresh the display. This parameter influences “Baud rate”, “Data output pause” and “Stream mode”.
Note
This item is selected automatically in the automatic response adjustment.
Decimal point (
pnt
)
The decimal point format can be selected.
Auto display-ON (
p-on
)
When the AC adapter is plugged in, the display is automatically turned on without the ON:OFF key
operation, to display the weighing mode. Used when the balance is built into an automated system.
30-minute warm up is necessary for accurate weighing.
Auto display-OFF (
poff
)
When the AC adapter is connected and no operation is performed (inactivity state) for 10 minutes,
the display is automatically turned off and the standby indicator illuminates.
Capacity indicator (
g5i
)
In the weighing mode, the indicator displays the weighing data relative to the weighing capacity in
percentage. (Zero =
0
%, maximum capacity =
100
%)
When the “Data memory (
data
)” parameter is set to “
1
” (to store unit mass in the counting mode),
“
2
” (to store the weighing data), “
4
” (to store comparator settings) or “
5
” (to store tare value), the
indicator displays the information stored in memory, such as the amount of memory data or data
number.
Accumulation function (
add
)
The accumulation function adds the weighing data, displays and outputs the total value. For details,
refer to “5-5 Accumulation Function.”
Display at start (
rng
)
When the weighing accuracy is not so strict, the smallest displayable weighing value can be turned off
without any key operation at weighing start. Useful when the balance is built into an automated system.
Содержание GP-61KS
Страница 8: ...6...
Страница 39: ...37 9 3 Details of the Function Table P...
Страница 40: ...38...
Страница 48: ...46 9 8 Data Format Examples...
Страница 49: ...47 Note When Pound Ounce is selected the data is output with the unit of ounce oz...
Страница 92: ...90 Separate Display Type GP 30KS 32KS 60KS 61KS 100KS Note The size in parentheses is for GP 100KS Unit mm...
Страница 97: ...95 MEMO...
Страница 98: ...96 MEMO...