68
Storing the tare value
To store a new tare value: Recall the stored tare value to be changed (“
t01
” to “
t20
”). Then, change
the recalled tare value using the digital input mode or the weighing input mode, and store the new
value.
Notes
When the RE-ZERO key is pressed with nothing placed on the weighing pan, zero is
displayed, The NET indicator does not illuminate.
“
t--
” appears when a tare operation is performed without using the tare value stored in
memory.
1 Press the MODE key to select a unit to be used for storage.
Note
While the data memory function is in use, unit selection using the MODE key is not available.
2 Set the “Data memory (
data
)” parameter to “
5
”.
If
“
Clr
” appears blinking in the upper left of the display, delete the data in memory as
described in “11-1 Notes on Using Data Memory”.
3 Press and hold the PRINT key until the balance enters the tare value confirmation mode. The
tare value last selected is displayed as shown below.
All the digits blinking
4 Select the tare number to be used, using the following keys.
RE-ZERO key
To increase the tare number by one.
MODE key
To decrease the tare number by one.
Each time the RE-ZERO key or MODE key is pressed, the displayed
value changes as follows: ....
↔
t03
↔
t04
↔
....
↔
t20
↔
t01
↔
....
CAL key
To cancel the operation and return to the weighing mode.
5 To change the selected tare value:
To use the weighing input mode, press the SAMPLE key to enter the weighing input mode. Go
to “Weighing input mode” on the next page.
To use the digital input mode, press the SAMPLE key, then press and hold the MODE key to
enter the digital input mode. Go to “Digital input mode” on the next page.
Note
Using the “PN:mm” command, the tare value can be recalled. (mm indicates a two-digit
numerical value 01-20, which corresponds to t01-t20.) The tare value recalled can be output
using the “?PT” command and can be changed using the “PT:” command.
Содержание GP-100K
Страница 8: ...6 ...
Страница 39: ...37 9 3 Details of the Function Table P ...
Страница 40: ...38 ...
Страница 48: ...46 9 8 Data Format Examples ...
Страница 49: ...47 Note When Pound Ounce is selected the data is output with the unit of ounce oz ...
Страница 92: ...90 Separate Display Type GP 30KS 32KS 60KS 61KS 100KS Note The size in parentheses is for GP 100KS Unit mm ...
Страница 97: ...95 MEMO ...
Страница 98: ...96 MEMO ...