51
Weighing input mode
Press the RE-ZERO key. The balance displays
0.0
g. Place a sample, with a mass that
corresponds to the upper limit value, on the pan. Press the PRINT key to store the upper limit
value. Remove the sample. The balance displays
Cp lo
.
9 With
Cp lo
displayed, press the PRINT key. The current setting of the lower limit value is
displayed with all of the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 10.
When the current setting is to be changed, press the RE-ZERO key. The balance is now in
the digital input mode. To use the weighing input mode, press and hold the MODE key.
Enter the lower limit value in the same way as described in step 8. Then, go to step 10.
10 Press the CAL key to exit the comparator function and return to the weighing mode.
Setting example 2
(Continuous comparison, including “near zero”, reference value and tolerance value)
Selecting a comparator mode
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed, then release the key.
2 Press the SAMPLE key several times to display
Cp fnc
.
3 Press the PRINT key.
4 Press the RE-ZERO key several times to display
Cp 4
.
5 Press the SAMPLE key to display
Cp in
.
6 Press the RE-ZERO key several times to display
Cp in 1
.
7 Press the PRINT key to store the selected mode.
Entering the reference and tolerance values
8
With
Cp ref
displayed, press the PRINT key. The current setting of the reference value is
displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 9.
When the current setting is to be changed, press the RE-ZERO key. The balance is now in
the digital input mode. To use the weighing input mode, press and hold the MODE key.
Digital input mode
Change the setting using the following keys.
SAMPLE key
To select the digit to change the value.
RE-ZERO key
To change the value of the digit selected.
MODE key
To switch the polarity.
PRINT key
To store the new setting and go to step 9.
CAL key
To cancel the new setting and go to step 9.
Weighing input mode
Press the RE-ZERO key. The balance displays
0.0
g. Place a sample, with a mass that
corresponds to the reference value, on the pan. Press the PRINT key to store the reference
value. Remove the sample and go to step 9.
Содержание GP-100K
Страница 8: ...6 ...
Страница 39: ...37 9 3 Details of the Function Table P ...
Страница 40: ...38 ...
Страница 48: ...46 9 8 Data Format Examples ...
Страница 49: ...47 Note When Pound Ounce is selected the data is output with the unit of ounce oz ...
Страница 92: ...90 Separate Display Type GP 30KS 32KS 60KS 61KS 100KS Note The size in parentheses is for GP 100KS Unit mm ...
Страница 97: ...95 MEMO ...
Страница 98: ...96 MEMO ...