22
Using the accumulation function
Use the keys below to operate the accumulation function.
MODE key:
Displays the weighing data and the total value alternately each time it is pressed.
Will not change the unit while the accumulation function is in use.
RE-ZERO key: Sets the display to zero while the weighing data is displayed.
Deletes the total value while the total value is displayed.
PRINT key
Outputs and adds the weighing data while the weighing data is displayed.
Outputs the total value while the total value is displayed.
1 Press the RE-ZERO key to set the display
to zero.
2 Place a sample on the pan. The weight
value is displayed.
3 Press the PRINT key. The weight value is
added to the total and output.
The accumulation number at the upper left
of the display increases by one.
4 Repeat steps 1-3 to accumulate more data.
5 Press the MODE key to display the total
value.
Outputting the value
6 Press the PRINT key to output the total
value.
7 Press the RE-ZERO key to delete the total
value.
Notes
The output format depends on the function
table setting.
While the accumulation function is in use,
the data memory function is not available.
To disable the accumulation funciton, set the
“Accumulation function (
add
)” parameter to “
0
”.
When the “Data number output (
d-no
)”
parameter is set to “
1
”, the accumulation
number will be output before the weighing
data.
Содержание GP-100K
Страница 8: ...6 ...
Страница 39: ...37 9 3 Details of the Function Table P ...
Страница 40: ...38 ...
Страница 48: ...46 9 8 Data Format Examples ...
Страница 49: ...47 Note When Pound Ounce is selected the data is output with the unit of ounce oz ...
Страница 92: ...90 Separate Display Type GP 30KS 32KS 60KS 61KS 100KS Note The size in parentheses is for GP 100KS Unit mm ...
Страница 97: ...95 MEMO ...
Страница 98: ...96 MEMO ...