40
9-4 Description of the Class “Environment, Display”
Condition (
Cond
)
Cond 0
This parameter is for sensitive response to the fluctuation of a weight value. Used for
powder target weighing, weighing a very light sample or when quick response weighing is
required.
After setting, the balance displays FAST.
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a weight value
from drifting due to vibration or drafts.
After setting, the balance displays SLOW.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the averaging time.
Stability band width (
5t-b
)
This item controls the width to regard a weight value as a stable value. When the fluctuation per second
is less than the parameter, the balance displays the stabilization indicator and outputs the data. The
parameter influences the “Auto print mode”
5t-b 0
This parameter is for sensitive response of the stabilization indicator. Used for exact
weighing.
5t-b 2
This parameter ignores slight fluctuation of a weight value. Used to prevent a weight value
from drifting due to vibration or drafts.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the stabilization
range.
Hold function (
Hold
) (Animal weighing mode)
This function is used to weigh a moving object such as an animal.
When the weight data is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the weighing pan,
the display returns to zero automatically.
This function is available only when the hold function parameter is set to “
1
”
(the animal mode indicator
ANIMAL illuminates) and any weighing unit other than the counting mode is selected.
The stabilization range and averaging time are set in “Condition (
Cond
)” and “Stability band width (
5t-b
)”.
Weighing range
FZ-120
i
WP
FZ-200
i
WP
FZ-300
i
WP
FX-120
i
WP
FX-200
i
WP
FX-300
i
WP
0.200 g or over
FZ-1200
i
WP
FZ-2000
i
WP
FZ-3000
i
WP
FX-1200
i
WP
FX-2000
i
WP
FX-3000
i
WP
2.00 g or over
Averaging time
Stabilization range
Cond 0
5t-b 0
Cond 1
5t-b 1
Cond 2
2 seconds
Faster
4 seconds
8 seconds More accurate
5t-b 2
Lesser
Greater
Содержание FZ-120iWP
Страница 7: ...5...
Страница 8: ...6...
Страница 47: ...45 9 7 Data Format Examples...
Страница 79: ......