31
5 Confirm that there is nothing on the pan and press the
key. The balance measures the zero point.
Do not allow vibration or drafts to affect the balance.
The balance displays the calibration weight value.
6 Place the displayed calibration weight on the
pan and press the
key. The balance
measures the calibration weight. Do not allow
vibration or drafts to affect the balance.
7 The balance displays
end
. Remove the
weight from the pan.
8 When the GLP output is set, the balance displays
glp
and outputs “Calibration Report”.
9 The balance will automatically return to the
weighing mode.
10 Place the calibration weight on the pan and
confirm that the value displayed is within
±
2 digits
of the specified value. If it is not within the range,
check the ambient conditions such as breeze and
vibration, also check the weighing pan. Then,
repeat steps 1 to 9.