20. SPECIFICATIONS
FZ-120
i
FZ-200
i
FZ-300
i
FZ-500
i
FZ-1200
i
FZ-2000
i
FZ-3000
i
FZ-5000
i
Weighing capacity
122 g
220 g
320 g
520 g
1220 g
2200 g
3200 g
5200 g
Maximum display
122.084 g 220.084 g 320.084 g 520.084 g 1220.84 g 2200.84 g 3200.84 g 5200.84 g
Minimum weighing value (1 digit)
0.001 g
0.01 g
Repeatability (Standard deviation)
0.001 g
0.01 g
Linearity
±
0.002 g
±
0.02 g
Stabilization time
(typical at FAST)
Approx. 1 second
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
±2 ppm/°C
±
0.010 g
±
0.10 g
±
0.15 g
Accuracy right after calibration
using the internal mass.
See notes below *
The value above is to the weighing capacity.
Internal mass
Yes
Clock and calendar function
Yes
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
5 times/second, 10 times/second or 20 times/second
Minimum unit mass
0.001 g
0.01 g
Counting
mode
Number of samples
5, 10, 25, 50 or 100 pieces
Minimum 100% reference mass
0.100 g
1.00 g
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
100 g
50 g
200 g
100 g
50 g
300 g
200 g
100 g
50 g
500 g
400 g
300 g
200 g
100 g
50 g
1000 g
500 g
2000 g
1000 g
500 g
3000 g
2000 g
1000 g
500 g
5000 g
4000 g
3000 g
2000 g
1000 g
500 g
Weighing pan diameter
130 mm
150 mm
Net weight
Approx. 3.0 kg
External dimensions
193 (W) x 262.5 (D) x 84.5 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface
RS-232C
* - Accuracy right after calibration using the internal mass, with the accessory breeze break
installed and in good ambient conditions (within the temperature range of 10
°
C to 30
°
C
(50
°
F to 86
°
F) with no abrupt changes in temperature or humidity, no drafts, no vibration
and no effect by magnetic fields or static electricity).
- With the FZ-
i
series balance, the environment and aging may cause internal mass value
error. Calibrate using an external mass and perform maintenance regularly.