background image

 

11

4.  DISPLAY SYMBOLS AND KEY OPERATION 

Key operation 

Key operation affects how the balance functions. The basic key operations are: 

  “Press and release the key immediately” or “Press the key”   

     = normal key operation during measurement 

  “Press and hold the key” 

             

 Display symbols 

 

 

 

 

 

 

 

 

 

 

 

 

Each key, when pressed or when pressed and held, functions as follows: 

Key 

When pressed 

When pressed and held 

 

Turns the display on or off. The standby indicator is displayed when the display is 
turned off. The weighing mode is enabled when the display is turned on. 
This key is available anytime. Pressing the key during operation will interrupt the 
operation and turn the display off. 

 

In the weighing mode, turns the minimum 
weighing value on or off. 
In the counting or percent mode, enters 
the sample storing mode. 

Enters the function table mode.   
Refer to “10. FUNCTION TABLE”. 

 

Switches the weighing units stored in the 
function table.   
Refer to “5. WEIGHING UNITS”.

 

Enters the response adjustment mode. 

 

Cancels the operation when performing 
settings. With the FZ-

i

 

series, enters the 

calibration mode using the internal mass. 
(One-touch calibration)

 

 

Enters the calibration mode.   

 

Outputs the weight data to a printer or 

personal computer using the RS-232C 

serial interface, depending on the 

function table settings. 
Confirms the operation when performing 

settings. 

No function at the factory setting 

By changing the function table: 
Outputs “Title block” and “End block” for 
GLP report. 

 

Sets the display to zero. 

Содержание FX-200i

Страница 1: ...i FX i SERIES FZ 120i FZ 200i FZ 300i FZ 500i FZ 1200i FZ 2000i FZ 3000i FZ 5000i FX 120i FX 200i FX 300i FX 500i FX 1200i FX 2000i FX 3000i FX 5000i Precision Balance 1WMPD4001863B INSTRUCTION MANUA...

Страница 2: ...ed into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject...

Страница 3: ...PCS 16 6 3 Percent Mode 18 6 4 Statistical Calculation Mode 19 6 5 Statistical Calculation Mode Example of Use 23 Adapting to the Environment 7 RESPONSE ADJUSTMENT 25 8 CALIBRATION 26 8 1 Calibration...

Страница 4: ...AND GLP REPORT 50 11 1 Setting the ID Number 50 11 2 GLP Report 51 12 UNDERHOOK 54 13 PROGRAMMABLE UNIT 55 RS 232C Serial Interface 14 RS 232C SERIAL INTERFACE 56 15 CONNECTION TO PERIPHERAL EQUIPMEN...

Страница 5: ...ne second to read a displayed value after a sample is placed on the pan has been achieved Multiple weighing units with most of the common units used around the world Standard RS 232C serial interface...

Страница 6: ...interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in...

Страница 7: ...5...

Страница 8: ...6...

Страница 9: ...1 2 3 2 UNPACKING THE BALANCE 2 1 Unpacking The balance is a precision instrument Unpack the balance carefully Keep the packing material to be used for transporting the balance in the future The packi...

Страница 10: ...0i Assemble the pan support and weighing pan on the balance as shown in the illustration on page 7 3 Adjust the leveling feet to level the balance Confirm it using the bubble spirit level 4 Confirm th...

Страница 11: ...ble should be solid and free from vibration drafts and as level as possible Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level Ensure a stable power source w...

Страница 12: ...lance interior free of dust and foreign materials The breeze break The FZ i series all and FX 120i 200i 300i 500i only is provided as a standard accessory An anti static treatment has been applied to...

Страница 13: ...d turn the display off In the weighing mode turns the minimum weighing value on or off In the counting or percent mode enters the sample storing mode Enters the function table mode Refer to 10 FUNCTIO...

Страница 14: ...rev Display Function table Storing mode Conversion factor 1 g Gram g g g 1 g Counting mode Pcs pcs pcs Percent mode Ounce Avoir Oz oz oz 28 349523125 g Pound Lb lb lb 453 59237 g Pound Ounce L OZ Loz...

Страница 15: ...660 0 00005 Tael China 3 90400 7 04000 10 24000 16 64000 0 00005 Tola India 10 4597 18 8618 27 4353 44 5823 0 0001 Messghal 26 0265 46 9335 68 2665 110 9330 0 0005 Capacity Unit FZ 1200i FX 1200i FZ 2...

Страница 16: ...order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when the displayed unit or mode is...

Страница 17: ...hile the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold separately such as a printer or a p...

Страница 18: ...g the counting mode 1 Press the MODE key to select 1pcs counting mode Storing a sample unit mass 2 Press the SAMPLE key to enter the sample unit mass storing mode Even in the storing mode pressing the...

Страница 19: ...s to be counted on the pan While the stabilization indicator is on pressing the PRINT key will output the weight value number of objects using the RS 232C serial interface Note Peripheral equipment th...

Страница 20: ...mass on the pan or in the container 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value Notes If the balance judges...

Страница 21: ...dard deviation and coefficient of variation are obtained by the equation below Standard deviation Standard deviation Coefficient of variation CV Average 5 4 1 Getting Started Switching to the Statisti...

Страница 22: ...described on page 22 and select the unit using the MODE key When the unit used for the statistical calculation mode is to be enabled upon power on select the unit in Unit Unit of the function table be...

Страница 23: ...he number of data is 1 the coefficient of variation is displayed as When the average is 0 the coefficient of variation is displayed as Statistical items are indicated on the upper left of the display...

Страница 24: ...t data and exclude it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data will be delete...

Страница 25: ...ase the key 2 Select the application function Press the SAMPLE key several times to display 1ap fnc Then press the PRINT key to display 1 apf 0 3 Change the application function parameter to 2 Press t...

Страница 26: ...0 00 g Storing the weight value of formula 1 The weight value data is output when the peripheral output equipment is connected 4 Weigh formula 2 and press the PRINT key The balance displays 0 00 g St...

Страница 27: ...SE is displayed And then release the key 2 Press the MODE key to select a rate of the response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT...

Страница 28: ...P using the RS 232C serial interface set GLP output info of Data output dout For details refer to 10 FUNCTION TABLE Calibration test is available only when GLP output info of Data output dout is set t...

Страница 29: ...Do not allow vibration or drafts to affect the balance 4 The balance displays end after calibration If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp an...

Страница 30: ...he internal mass value Refer to 9 FUNCTION SWITCH AND INITIALIZATION and set the internal mass value correction switch to 1 3 In the weighing mode press and hold the SAMPLE key to display ba5fnc 4 Pre...

Страница 31: ...INITIALIZATION and set the internal mass value correction switch to 1 2 In the the weighing mode press an d hold the SAMPLE key to display ba5fnc 3 Press the SAMPLE key several times until C5 in 21 i...

Страница 32: ...z If you use the calibration weight value stored in the balance go to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of weight selection mod...

Страница 33: ...ght Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays 1 glp and outputs Calibration Rep...

Страница 34: ...usable weights is shown o n page 26 press the SAMPLE key and go to step 4 z If you use the target weight value stored in the balance go to step 5 4 Specify the target weight value as follows SAMPLE ke...

Страница 35: ...et weight value 6 Place the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect...

Страница 36: ...calibration 0 To inhibit calibration using the internal mass 1 To permit calibration using the internal mass Calibration using an external weight 0 To inhibit calibration using an external weight 1 T...

Страница 37: ...once again 9 2 Initializing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 reference mass...

Страница 38: ...e A for Data output mode and 100 digits for Auto print difference 10 2 Display and Keys Display Key Description The symbol indicates that the parameter displayed is in effect When pressed and held in...

Страница 39: ...y ON 1 ON Turns on the weighing mode display when the AC adapter is connected 0 OFF poff Auto display OFF 1 ON 10 minutes Turns off the display after 10 minutes of inactivity 0 Displays rng Display at...

Страница 40: ...nt polarity 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap b Auto print difference 2 1000 digits Difference between reference value and displayed value 0 Every measurement 1 2 second...

Страница 41: ...efer to 5 WEIGHING UNITS id ID number Refer to 11 ID NUMBER AND GLP REPORT 0 Normal weighing mode 1 Capacity indicator apf Application function 2 Statistical calculation mode 0 Number of data sum 1 Nu...

Страница 42: ...ue Used to prevent a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item is used to set the stabilization range Hold function Hold Animal weighing...

Страница 43: ...splay refresh rate 5pd Period to refresh the display This parameter influences Stream mode along with Baud rate and Data output pause Note This item is selected automatically when the response rate is...

Страница 44: ...added with ar d set to 1 to adjust zero after the data is output Mode B Required setting dout prt 2 Auto print mode B reference laststablevalue dout ap p Auto print polarity dout ap b Auto print diffe...

Страница 45: ...at If an AD 8121B is used set the printer to MODE 1 or 2 This format consists of fifteen characters excluding the terminator A header of two characters indicates the balance condition The polarity sig...

Страница 46: ...tes the balance condition The polarity sign is used only for negative data The weight data uses spaces in place of the leading zeros The character length of this format changes dependent upon the unit...

Страница 47: ...T NU E H Unstable LF CR LF CR LF CR LF CR A D DP KF MT NU CR LF CR LF LF Negative error Overload CR LF CR CR A D DP KF MT NU L LF U S 0 0 1 8 3 9 6 g U S 1 8 3 9 6 g 1 8 3 9 6 0 0 1 8 3 9 6 1 8 3 9 6...

Страница 48: ...pacity indicator apf 1 In the weighing mode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mo...

Страница 49: ...ll the digits blinking When the date is correct and the operation is to be finished press the CAL key and go to step 8 When the date is correct and the time is to be confirmed press the SAMPLE key and...

Страница 50: ...t the time in a 24 hour format using the following keys SAMPLE key To select the digits to change the value The selected digits blink RE ZERO key To increase the value by one MODE key To decrease the...

Страница 51: ...per limit value 6 With Cp Hi displayed press the PRINT key The current setting of the upper limit value is displayed with all the digits blinking When the current setting is not to be changed press th...

Страница 52: ...n output the following reports for GLP Calibration report of the calibration using the internal mass Calibration due to changes in temperature and one touch calibration Calibration report of the calib...

Страница 53: ...e are not correct adjust the AD 8121B clock and calendar FX i series or use Cl adj of the function table to adjust the time and date FZ i series Calibration report using an internal weight When the se...

Страница 54: ...IGHT 2000 00 g SIGNATURE TERM CR LF CR Space ASCII 20h Terminator or Carriage return ASCII 0Dh Line feed ASCII 0Ah CR LF AD 8121 format General data format Manufacturer Model Serial number ID number D...

Страница 55: ...data Time Signature End block AD 8121 format Manufacturer Model Serial number ID number Date Time Title block General data format A D TERM MODEL FX 3000i TERM S N 01234567 TERM ID ABCDEFG TERM DATE T...

Страница 56: ...is revealed by removing the plastic cap on the bottom of the balance Use the underhook as shown below Caution Do not apply excessive force to the underhook When not in use attach the plastic cap to p...

Страница 57: ...set the coefficient Confirming the coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and go to step 6 When it is to be ch...

Страница 58: ...Parity Even Odd Data bits 7 bits None Data bits 8 bits Stop bit 1 bit Code ASCII D Sub 9 pin assignments Pin No Signal name Direction Description 1 No connection 2 TXD Output Transmit data 3 RXD Input...

Страница 59: ...0 0 0 0 1 2 bp5 Baud rate 2 2 2 2 btpr Data bit parity bit 0 0 0 0 Crlf Terminator 0 0 0 0 5if Serial interface type Data format 0 0 0 1 1 Set appropriate parameters for ap p Auto print polarity and...

Страница 60: ...om and RsKey RsCom Can transmit commands to control the balance Can make bi directional communication between the balance and a personal computer using the RS 232C serial interface Can display or stor...

Страница 61: ...nds commands such as re zero or send weight data to the balance and controls the balance 3 Printing the balance GLP report using your printer The balance GLP report can be printed using a printer conn...

Страница 62: ...ommands the S and E SCP commands behave the same Commands to control the balance CAL Same as the CAL key EXC Calibration using an external weight OFF Turns the display off ON Turns the display on P Sa...

Страница 63: ...the balance and can not process it the balance transmits an error code EC Exx When the balance receives a command to control the balance and can process it the balance transmits the acknowledge code...

Страница 64: ...Personal computer Time Balance CAL command Zero display Confirmation of command receipt Confirmation of command receipt Confirmation of command receipt Command completion Starts inputting data Zero d...

Страница 65: ...lance has two functions Data output dout and Serial interface 5if Set each function as necessary PT command Tare of 100g PT command LF CR T P PT 00100 00 gCR LF P T 1 0 0 Tare value display AKCR LF g...

Страница 66: ...or weighing method is proper Operating environment Is the weighing table solid enough especially for the FZ 120i 200i 300i 500i and FX 120i 200i 300i 500i the balances with a minimum weighing value o...

Страница 67: ...ight value Press the CAL key to return to the weighing mode EC E21 Calibration weight error The calibration weight is too light Confirm that the weighing pan is properly installed Confirm the calibrat...

Страница 68: ...acter of a command within the time limit of one second Confirm the communication EC E04 Excess characters error The balance received excessive characters in a command Confirm the command EC E06 Format...

Страница 69: ...N The WinCT Plus data communication software is provided as a standard accessory and can perform the following Acquire data from multiple balances connected to a LAN Control these balances with comman...

Страница 70: ...ze weighing errors due to static electricity on the material The AD 1683 is direct current static eliminator The ions generated produce no breeze and are effective over a long distance Therefore the b...

Страница 71: ...interface to display the weight data and to remotely control the balance Analog output and comparator output available as an option to be installed AD 8121B Printer Compact dot matrix printer Statist...

Страница 72: ...ces Minimum100 referencemass 0 100 g 1 00 g Percent mode Minimum100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 100 g 50 g 200 g 100 g 50 g 300 g 200 g 100 g 50...

Страница 73: ...Display refresh rate 5 times second 10 times second or 20 times second Minimum unit mass 0 001 g 0 01 g Counting mode Number of samples 5 10 25 50 or 100 pieces Minimum100 referencemass 0 100 g 1 00 g...

Страница 74: ...accessory Unit mm 1 Width of the opening when a clear plate is removed 2 Inside dimension 3 Weighing pan diameter 4 Height from the weighing pan up to the cap of the breeze break Inside dimension 5 He...

Страница 75: ...r which is not included in the weight data Mode Balance operational function Re zero To set the display to zero GLP Good Laboratory Practice Repeatability Variation in measured values obtained when th...

Страница 76: ...modeA 38 42 Auto print mode B 38 42 Auto print polarity 38 AVE Average 21 Average 19 21 39 AX USB 25P EX USB converter 68 B ba5fnc Environment display 14 37 Baud rate 39 56 57 beep Beep 37 bep HI buz...

Страница 77: ...09 67 FXi 10 67 FXi 11 68 FXi 12 68 FXi 15 68 FXi 31 68 G GLP 50 GLP output 26 31 32 38 51 Grounding terminal 7 H Hold Hold function 37 40 I ID number 50 info GLP output 38 Inhibit 34 Initializing the...

Страница 78: ...ity drift 70 71 5if Serial interface 39 SLOW 25 Small breeze break 67 5pd Display refresh rate 37 41 Stability band width 40 Stabilization indicator 11 Stabilization time 70 71 Standard deviation 19 2...

Отзывы: