37
10-3 Details of the Function Table
Class
Item and Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
With “
Hold 1
”, sets the averaging
time.
0
Stable range is ±1 digit
1
5t-b
Stability band width
2
Stable range is ±3 digits
The stabilization indicator illuminates
when the display fluctuation is
within the range. With “
Hold 1
”,
sets the stabilization range.
0
OFF
Hold
Hold function
1
ON
Holds the display when stable in animal
mode. With “
Hold 1
”, ANIMAL turns on.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
1
10 times/second
5pd
Display refresh rate
2
20 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
0
OFF
poff
Auto display-OFF
1
ON (10 minutes)
Turns off the display after 10 minutes
of inactivity.
0
Displays
rng
Display at start
1
Does not display
Select whether or not to display the
minimum weighing value at weighing
start.
0
Does not sound
ba5fnc
Environment
Display
beep
Beep
1
Sounds
Select whether or not to sound the
beep when operating on keys.
Cl adj
*
Clock
Refer to “10-9 Clock and calendar
function”.
Confirms and sets the time and
date. The time and date are added
to output data.
0
No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
3
Continuous comparison, excluding "near zero"
Cp
Comparator mode
4
Continuous comparison, including "near zero"
0
OFF
bep-
LO buzzer
1
ON
0
OFF
bep-
OK buzzer
1
ON
0
OFF
Cp fnc
Comparator
bep-
HI buzzer
1
ON
: Factory settings. Digit is a unit of minimum weighing value
*
:
Only for
the
FZ-
i
series