![A&D FX-200i Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/aandd/fx-200i/fx-200i_instruction-manual_3080286054.webp)
52
Calibration report using an external weight
When the setting is “
info 1
”:
When the setting is “
info 2
”:
Calibration test report using an external weight
(Calibration test does not perform calibration.)
When the setting is “
info 1
”:
When the setting is “
info 2
”:
AD-8121 format
General data format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration
weight
Signature
A & D<TERM>
M ODEL F X- 300 0i <TERM>
S /N 0 12 345 67 <TERM>
I D AB CDE FG <TERM>
D ATE<TERM>
<TERM> *
T IME<TERM>
<TERM> *
C ALIB RA TE D( EXT .)<TERM>
CAL.WEIGHT<TERM>
+2 00 0. 00 g <TERM>
S IGNA TU RE <TERM>
<TERM>
<TERM>
- - -- - - -- - - - -- - - -<TERM>
<TERM>
<TERM>
A &
D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2008/12/31
TIME 12:23:34
CALIBRATED(EXT.)
CAL.WEIGHT
+2000.00 g
SIGNATURE
<TERM>
CR LF CR
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
CR
LF
AD-8121 format
General data format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test
type
Zero point value
Target weight
value
Target weight
Signature
A & D<TERM>
M ODEL FX -3 00 0i <TERM>
S /N 01 23 45 67 <TERM>
I D A BC DE FG <TERM>
D ATE<TERM>
<TERM> *
T IME<TERM>
<TERM> *
C AL.T ES T(E XT .)<TERM>
ACTUAL<TERM>
0 .0 0 g <TERM>
+1 999 .9 9 g <TERM>
TARGET<TERM>
+2 000 .0 0 g <TERM>
S IGNA TU RE<TERM>
<TERM>
<TERM>
- - - - - - - - - - - - - - - - <TERM>
<TERM>
<TERM>
A &
D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2008/12/31
TIME 12:23:34
CAL.TEST(EXT.)
ACTUAL
0.00 g
+1999.99 g
TARGET
+2000.00 g
SIGNATURE
<TERM>
CR LF CR
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
CR
LF
With FZ-
i
series, the time and
date are added to output data.
With FZ-
i
series, the time and
date are added to output data.