![A&D EK-i Series Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/aandd/ek-i-series/ek-i-series_instruction-manual_3210000031.webp)
29
1. Press and hold the calibration (CAL) switch.
CC
appears after
Cal
. Release the
switch when
CC
is displayed.
2.
CC 0
is displayed.
3. Press the SAMPLE key and change the weight
value using the following keys.
SAMPLE key
To select the digit blinking to
change.
RE-ZERO key
To set the value of the digit
selected.
PRINT key
To store the value and return
to step 2.
4. At step 2, press the PRINT key. The zero point is
weighed and the weighed value is displayed for a
few seconds.
5. Place a weight, of the same value as displayed, on
the pan. Press the PRINT key to weigh it. The
weighed value is displayed for a few seconds.
2000.0
Calibration
weight
Weighing
pan
02000.0
CC 0