![A&D EK-i Series Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/aandd/ek-i-series/ek-i-series_instruction-manual_3210000025.webp)
23
q
There are four types of headers:
ST : Stable weighing data (including % data)
QT : Stable counting data
US : Unstable weighing data (including count and %)
OL : Out of weighing range (Over)
q
The data is normally 9 digits including a decimal point and a sign.
q
There are 11 types of units:
g : Weighing data “gram”
P C : Counting data “pcs”
% : Percentage data “%”
o z : Weighing data “decimal ounce”
l b : Weighing data “decimal pound”
o z t : Weighing data “troy ounce”
c t : Weighing data “carat”
m o m : Weighing data “momme”
d w t : Weighing data “penny weight”
G N : Weighing data “grain”
t l : Weighing data “tael”
q
The terminator is always C
R
L
F
.
q
Example of output data:
Weighing data “gram”
S T , + 0 0 1 2 3 4 . 5
g
C
R
L
F
Counting data
Q T , + 0 0 0 1 2 3 4 5
P C
C
R
L
F
Percentage data
S T , + 0 0 0 1 2 3 . 4
%
C
R
L
F
Out of range “gram” (+)
O L , + 9 9 9 9 9 9 . 9
g
C
R
L
F
Out of range “pcs” (-)
O L , - 9 9 9 9 9 9 9 9
P C
C
R
L
F
9-3. Data output mode
Stream mode
Set the function “
pr t 0
”.
The balance outputs the current display data. The data-update rate is approximately
10 times per second. This rate is the same as the display-update.
The balance does not output data while it is in the setting mode.
Key mode
Set the function “
pr t 1
,
2
or
3
”.
When the PRINT key is pressed while the weighing data is stable (the stability mark
is on), the balance transmits the data. When the data is transmitted, the display will
blink one time.