15
5. At step 3, pressing the PRINT key weighs the zero-
point value. Do not touch the pan during weighing.
6. Place the calibration weight with the same value as
displayed on the pan. Press the PRINT key to
weigh it. Do not touch the pan during weighing.
7.
end
appears.
Remove the weight from the pan, and press the CAL
switch or MODE key to return to the weighing mode.
q
q
Note
The value set in step 4 is stored in memory even after the power is
switched off.
If the balance is to be moved to other places, set the gravity
acceleration value of the area where the calibration using a weight is
to be done, and calibrate the balance according to the procedure
above. See the next section to set the value.
7-2. Gravity acceleration correction