![A&D EK-i Series Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/aandd/ek-i-series/ek-i-series_instruction-manual_3210000026.webp)
24
Auto-print mode A
Set the function “
pr t 2
”.
The balance transmits the weighing data when the display is stable (the stability mark
is on) and the data is greater than +4d (of the lowest range for EW-
i
series).
The next output can be obtained after the display returns below +4d.
Auto-print mode B
Set the function “
pr t 3
”.
The balance transmits the weighing data when the display is stable (the stability mark
is on) and the data is greater than +4d (of the lowest range for EW-
i
series) or less
than -4d.
The next output can be obtained after the display returns between -4d and +4d.
9-4. Command mode
In the command mode, the balance is controlled by commands that come from the
personal computer and so on.
Command list
q
Command to request the current weighing data.
Command
Q C
R
L
F
Reply
S T , + 0 0 1 2 3 4 . 5
g C
R
L
F
q
Command to zero or tare the balance (same as the RE-ZERO key).
Command
Z C
R
L
F
Reply
Z C
R
L
F
q
Command to change the weighing units (same as the MODE key).
Command
U C
R
L
F
Reply
U C
R
L
F
10. ID NUMBER AND GLP
The ID number is used to identify the balance when Good Laboratory Practice (GLP)
is used. The following GLP data is transmitted to an AD-8121 printer or a computer
using the RS-232C interface.