22
Detailed information on Holding Register 2
(
only for AD-4212C/AD-4212D
)
The following items change only when connected to the the AD-4212C / AD-4212D.
When connected to another measuring instrument, they will be the initial values.
■
Response characteristics (Holding Register: 40 2015 to 2016)
⋆1.
FAST / MID / SLOW can be switched by writing “1” to Coil 00 2006.
⋆2.
AD-4212C with fine response characteristics.
(
Refer to WinCT-AD4212C additional instruction manual for setting method.
)
■
Calibration status (Holding Register: 40 2017 to 2018)
Value
Response characteristic status
0x00000000
Default value
0x00000001
Waiting for zero point weighing
0x00000002
Weighing zero point
0x00000003
Waiting for weighing of calibration weight
0x00000004
Weighing calibration weight
0x00000005
Calibration completed
0x00000006
Calibration error
■
Calibration weight value (Holding Register: 40 2019 to 2020)
When executing calibration (sensitivity adjustment) with your own weight, the value of the calibration weight can be
checked.
‐
The unit is gram and the decimal point is none.
‐
Only when “Waiting weighing of calibration weight” is the value of the calibration weight to be used.
The initial value is “0xFFFFFFFF”. When returning to the weighing mode, it will return to the initial value.
Calibration weight value
to be used
Data Address
(Holding Register)
Data
Description
100g
40
2019
0x0064
Calibration weight value 0x00000064
⇒
100
2020
0x0000
Value
Response characteristic status
⋆
1
0x00000000
Default value
0x00000001
FAST
0x00000002
MID
0x00000003
SLOW
0x00000004
User setting
⋆
2