79
8.14. Connection
8.14.1. Serial
mode
Set the serial communication mode.
Note:
To perform serial communication, the checkweigher must be connected to the serial interface.
For details, refer to “7.4. Connecting to the Serial Interface.”
Each mode is as follows:
1. Disable
Select to disable the serial communication.
2. Cyclic
printing
Set when a printer for cyclic printing is connected.
When “Cyclic printing” is selected, data for cyclic printing will be output through serial communication.
3. A&D
format
Data is output in A&D format as shown below.
A&D format
The weighing data is fixed to 15 characters excluding terminators <CR><LF>.
The first two characters are the header to indicate the status of the weighing value.
The polarity sign is placed before the data. If the data is zero, the plus sign is used.
Output example
1 2 3 4 5 6 7 8 9 10
11
12
13
14 15
S T , + 0 0 0 1 2 . 3 4 5 g
Stable
S T , + 0 0 0 1 2 . 3 4 5 g
Unstable
U S , + 0 0 0 0 5 . 4 3 2 g
Overloaded (positive)
O L , + 9 9 9 9 9 9 9 E + 1 9
Overloaded (negative)
O L , - 9 9 9 9 9 9 9 E + 1 9
Header
Data
Unit