AD-4411
9.3. EtherCAT (AD-4411-ECT)
9.3.1. Part names
Name Function
Description
RUN Communication status LED
OFF
Power-Off or Init state
Green, ON
Operational state
Green, blinking
Pre-Operational state
Green, single blink
Safe-Operational state
ERR Error status LED
OFF
Power-Off or no error
Red, blinking
Communication setting error
Red, single blink
Communication data error
Red, double blinks
Watchdog timeout
9.3.2. Communication specifications
Communication is performed according to the following communication specifications.
Note that communication is possible only in measurement mode.
Communication standard
EtherCAT (Sync mode is supported only for Free Run.)
Transmission speed
100 Mbps
Connector
RJ-45 (P1 = IN, P2 = OUT)
Cable specifications
Shielded twisted pair (STP) cable, straight / crossover, CAT 5e or above
Configuration file
ESI file (Please download the configuration file from A&D’s website.)
9.3.3. Communication settings
When using the Explicit Device ID, set it with the Ect function.
If "0: Not used (default)", the PLC settings are valid.
Ect F
Setting item
Setting value
Default
01
Explicit Device ID
0: Not used
1~65535
0
RUN
ERR
Ethernet port
OUT
IN