8. Basic functions
After AD-4411 is turned on, all segments of the display turn on / off to check the display. Then AD-4411 enters
measurement mode and starts measuring.
Refer to "14 Setting functions" for function setting methods.
8.1. Main display
The following content is displayed in measurement mode.
・
Measurement value.
・
Positive overload.
The gross value exceeds the maximum capacity by +8d.
・
Negative overload.
The gross value meets the negative overload condition (CALF15).
(Default = 0: Gross < - maximum ca 8d)
8.2. Power-on zero
After power-on, the gross value is set to zero automatically when the gross value is within the maximum
capacity ± 10% and stable.
Set power- on zero (CALF14) to 1: Enable (default = 0: Disable) to use this function.
If the condition is not met, "------" is displayed until the condition is met.
Press the [MODE/EXC] key to skip power-on zero and enter measurement mode.