3.3. General
3.3.1. Interface
Standard serial output
Output for communication with A&D peripheral devices (20
mA current loop)
CC-Link interface
CC-Link ver.1.10 remote device station
3.3.2. Measuring functions
Zero adjustment
•
Sets the gross weight to zero by pressing the ZERO key or
a command from the CC-Link master station.
•
Selection of disable or enable for the operation when
unstable.
•
The zero value is battery backed up.
•
Zero adjustable range
Can be set optionally in the range of 1 to 100% of the
weighing capacity.
Zero tracking
•
Tracks the weight drift around the zero point to maintain
zero.
•
Zero tracking time
0.0 to 5.0 seconds (Can be set optionally within the range.)
•
Zero tracking band
0.0 to 9.9 d (Can be set optionally within the range.)
Tare
•
Sets the net weight to zero by pressing the TARE key or a
command from the CC-Link master station.
•
Selection of disable or enable for the operation when
unstable or negative.
•
The tare weight is battery backed up.
•
Tare range: Gross weight
≤
Weighing capacity
Stability detection
•
Turns ON the stabilization indicator when the variables of
the weight values per sampling are within the set band in
the set time.
•
Stability can be confirmed using the CC-Link.
•
Stability detection time
0.0 to 9.9 seconds (Can be set optionally within the range.)
•
Stability detection band
0 to 9 d (Can be set optionally within the range.)
Digital filter
Cutoff frequency range: 0.07 to 11 Hz