6.1.5.
display
Assign a function to the
display in the general functions.
Functions related to the
display
z
fncf04
: Assigns a function to the
display from the functions below:
0
: None
1
: Zero tracking in progress
2
: Alarm (Zero range setting error, over, low battery)
3
: F key status (Read the description below.)
F key status
z
Enables when alternate switch or momentary switch is selected for
fncf02
.
The display turns ON when the F key is ON and turns OFF when the F key is OFF.
6.1.6. Memory backup
The indicator has two kinds of memory.
EEPROM
Memory used to store important data, without power
supplied, that the occurrence of re-writing is seldom.
Data example: Calibration data, Function data
Battery backup RAM
Memory used to store temporary data that the occurrence
of re-writing is often.
Data example: Zero value, Tare weight
Memory backup
Data Method
Description
Calibration data
Function data
EEPROM Maintained
regardless of battery condition.
Zero value
Tare weight
Battery
backup RAM
The life of the battery is approximately 10 years
at 25°C with the power disconnected.