10.2.4. Checking the CC-Link
With
Chc CC
displayed, press the ENTER key to enter the CC-Link check mode.
Press the
∧
or
∨
key to change the address.
Press the CAL key to transmit the request flag of initialization.
Address
Remote I/O
Buffer memory
05 to 0C
RY0000 to RY007F
160 to 167
0D to 1C
RWw0000 to RWw000F
1E0 to 1EF
25 to 2C
RX0000 to RX007F 0E0
to
0E7
2D to 3C
RWr0000 to RWr000F
2E0 to 2EF
10.2.5. Checking the testing terminal
The
Chc r5
display is for checking the testing terminal. Do not use it.
10.2.6. Checking the A/D (load cell)
With
Chc ad
displayed, press the ENTER key to enter the A/D check mode. The load cell
output value is displayed in mV/V.
The example below is when the internal count is 1.23456 mV/V.
If the internal count exceeds ±7mV/V, the load cell may be damaged or the load cell may
be connected incorrectly. Refer to “10.4. Checking the Load Cell Connection Using the
Digital Multimeter” to check the connection.
10.2.7. Checking the internal count
With
Chc in
displayed, press the ENTER key to enter the internal count check mode.
The internal count value (display value times 10) is displayed.
The example below is when the internal count is 123.
1 . 2 3 4 5 6
1 2 3 .
8 8 . 8 8 8 8
Remote READY flag
Error status flag
Reply flag of initial data setting
Request flag of initialization
CPU normal operation
Link status
2 digits: Address (00 to 3F)
4 digits: Data (0000 to FFFF)