![A&D AD-4408C Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/aandd/ad-4408c/ad-4408c_instruction-manual_446087049.webp)
Page
47
AD-4408C Weighing Indicator
Resuming from suspended modes
When the AD-4408C is in a mode that suspends weighing, such as calibration or OFF
mode, the remote READY flag (RX007B) becomes inactive because a correct weight
value can not be output.
To resume from this condition, take the same steps as described in “When the
AD-4408C is turned on” above.
Requesting initialization from the master station
When requesting the initial data setting to the AD-4408C from the master station, turn
the request flag of initial data setting (RY0079) ON while the remote READY flag
(RX007B) is active.
The AD-4408C turns the remote READY flag (RX007B) OFF and performs initial data
settings.
When initial data settings are complete, the reply flag of initial data setting (RX0079) is
turned ON.
Turn OFF the request flag of initial data setting (RY0079) in the master station.
Fig.11 Performance of request flag of initial data setting
Remote READY flag (RX007B)
Request flag of initial data setting (RY0079)
Reply flag of initial data setting (RX0079)
When the remote READY flag
(RX007B) is ON, the request flag
of initial data setting (RY0079) is
turned ON.
When the reply flag of initial data
setting (RX0079) is ON, the
request flag of initial data setting
(RY0079) is turned OFF.
When the reply flag of initial data setting
(RX0079) is OFF, the remote READY
flag (RX007B) is turned ON again.