Page
42
AD-4329A Weighing Indicator
11. Standard Serial Data Output (SER.OUT)
The standard serial data output can be used to output the data to an A&D printer and
a display unit.
Set
the
f-30
~
f-34
of the F-Functions to use this output.
When using a non-buffered printer like the AD-8117, AD-8117A and AD-8121 etc, the
printer can not receive data while printing. Therfore, set to
f-33 1
for the manual print
mode or auto print mode.
Pin No.
Signal
3 pin
5 pin
Current loop
Other pins Not used
Case Shield
Inside of AD-4329A
Related Functions
F-Function
Description
f-30 0
Displayed data
f-30 1
Gross data
f-30 2
Net data
f-30 3
Tare value
Output Data
f-30 4
Gross data / Net data / Tare value
Output mode
F-Function
Description
Stream mode
f-31 0
Description of "Stream Mode" at page 31
Auto print mode
f-30 1
Description of "Auto Print Mode" at page 31
Manual print mode
f-30 2
When pressing the
key each time, the data is output once.
Auto-print condition
F-Function
Description
Positive data only
f-32 0
"Inhibition region of output" +5d.
+5d < "permission region of output".
Positive and
negative data
f-32 1
-5d "inhibition region of output" +5d.
"Permission region of output" < -5d, +5d < "permission region of output".
F-Function
Description
f-33 0
Not used
Delay for continual
data
f-33 1
Approximately 2 seconds
Baud rate
f-34
600 bps, 1200 bps, 2400 bps
CF-Function Description
Cf-07
Data output of tare
Cf-08
Output on over load and unstable state. Use or not used
Cf-09
Format of Header
Selection of
TR
,
PT
,
GS
,
NT
,
T
,
G
and
N
Cf-10
Figure number of unit, decimal point
Two figures or three figures, dot or comma
11. Standard Serial Data Output
Содержание AD-4329A
Страница 1: ...Weighing Indicator 1WMPD4002801 UHYLVHG...
Страница 54: ...Page 52 AD 4329A Weighing Indicator MEMO...
Страница 55: ......